Load 1 (1/5 max), Load 2 (1/3 max), Load 3 (2/3 max) – KERN & SOHN CFS 50K-3 User Manual

Page 66: Load 4 (max), Load 0, Load 3 (2/3 max, Load 1 (1/3 max), Load 2 (2/3 max), Load 3 (max)

Advertising
background image

Table 2: Required adjustment weights for interconnected bulk scales

1. Counting systems with reference scales KERN CFS 300-3, CFS 3K-5

6 kg

15 kg

30 kg

60 kg

150 kg

300 kg

600 kg

1500 kg

3000 kg

load 1

(1/5 Max)

1 kg

3 kg

5 kg

10 kg

30 kg

60 kg

100 kg

300 kg

600 kg

load 2

(1/3 Max)

2 kg

5 kg

10 kg

20 kg

50 kg

100 kg

200 kg

500 kg

1000 kg

load 3

(2/3 Max)

4 kg

10 kg

20 kg

40 kg

100 kg

200 kg

400 kg

1000 kg

2000 kg

load 4

(max)

6 kg

15 kg

30 kg

60 kg

150 kg

300 kg

600 kg

1500 kg

3000 kg

load 0

0 kg

0 kg

0 kg

0 kg

0 kg

0 kg

0 kg

0 kg

0 kg

load 4

(max)

6 kg

15 kg

30 kg

60 kg

150 kg

300 kg

600 kg

1500 kg

3000 kg

load 3

(2/3 Max

4 kg

10 kg

20 kg

40 kg

100 kg

200 kg

400 kg

1000 kg

2000 kg

load 2

(1/3 Max)

2 kg

5 kg

10 kg

20 kg

50 kg

100 kg

200 kg

500 kg

1000 kg

load 1

(1/5 Max)

1 kg

3 kg

6 kg

10 kg

30 kg

60 kg

100 kg

300 kg

600 kg

2. Counting systems with reference scale KERN CFS 50K-3

150 kg

300 kg

600 kg

1500 kg

3000 kg

load 1

(1/3 Max)

50kg

100kg

200kg

500kg

1000kg

load 2

(2/3 Max)

100kg

200kg

400kg

1000kg

2000kg

load 3

(max)

150kg

300kg

600kg

1500kg

3000kg


For counting systems with reference scale CFS 6K0.1, CFS 15K0.5 or
CFS 30k0.5, linearization of the bulk scales is impossible.