 Use the numeric keys to input password „9999“:

Confirm input by

.

„Pin“ „----„

 Select bulk material or reference balance via

.

The appeared [

] displays the current balance.


When using as counting system, the bulk material
balance as well as the reference balance must be
adjusted. The adjustment process must be carried out
on both balances.

 Acknowledge with

.

„tECH“ „LoCAL“

„tECH“ „rEmotE“

 Using

select the weighing unit [kg or lb] which will

be used for linearization. The appeared [

] displays the

current weighing unit. Acknowledge with

.

„tECH“ „Unit“

 Ensure that there are no objects on the weighing pan.

Wait for stability display (indicator [

] atop

appears), then press .

 Carefully place first adjustment weight into the center of

the weighing plate on display „LoAd 1“.

Wait for stability display, then press

.

 Carefully place second adjustment weight into the

center of the weighing plate on display „LoAd 2“.

Wait for stability display, then press

.

 When „LoAd 3“ is displayed, place the third adjustment

weight carefully in the centre of the weighing plate.

Wait for stability display, then press

.

 After the adjustment the balance will carry out a self-

test. The balance returns automatically into weighing
mode. In case of an adjustment error or incorrect
adjusting weight the display will show an error message
(

), repeat adjustment process.