13 Configuration of bulk scale

 Changes may only be carried out by trained specialized personnel.


Ex factory, the scales KERN CFS respectively counting systems KERN CCS have
been programmed such that no changes are generally necessary.
But if there are special conditions of use or if as bulk material balance an other
weighing bridge (not preconfigured by KERN) is connected, in the menu block
"

" the required settings can be made.


Technical data

Supply voltage:

5 VDC

Max. signal voltage

0-20 mV

Zeroing range

0-5 mv

Sensitivity

> 0.02

μv

Resistance parameter

87 O Min., 4 x 350 Ω load cell

Connection

4 poles

Max. cable length

6 meter

Connection plug

9 pin d-subminiature bushing


Navigation in the menu:

 With

or

(model CFS 50K-3), the individual menu items may be

selected one by one.

 Confirm selected menu item with

or

(model CFS 50K-3). The current

setting will be displayed.

 With

bzw.

(model CFS 50K-3) switch over to the available settings.

 Either save with

bzw.

(model CFS 50K-3) or reject with

or

(model CFS 50K-3).


CFS/CCS-BA-e-1422