14 Adjustment

• Provide required adjustment weight, cf. chapter 1.

The adjustment weight to be used depends on the capacity of the weight /
counting system. Implement adjustment as close to the maximum load as
possible. Info about test weights can be found on the Internet at:
http://www.kern-sohn.com

• Observe stable environmental conditions. A warming up time (see chapter 1)

is required for stabilization.

• In order to avoid errors at the piece quantity determination, both balances

must be adjusted with the same acceleration due to gravity.
In case of non-compliance counting errors will result!

14.1 Models CFS 300-3, CFS 3K-5

Operation

Display

 Switch-on balance and during the selftest press

.

„Pin“

 Use the numeric keys to enter password:

Either input four times zero „0000“ as a standard
password, or input the user-defined password (input cf.
parameter „Pin“ chapter 13).

 Confirm input by

.

„Pin“ „----„

 Select bulk material or reference balance via

.

The appeared [

] displays the current balance.


When using as counting system, the bulk material
balance as well as the reference balance must be
adjusted. The adjustment process must be carried out on
both balances

„tECH“ „LoCAL“

„tECH“ „rEmotE“

 If necessary, at balance zero display using

select

the weighing unit [g / kg or lb], which shall be used for
adjustment. The appeared [

] displays the current

weighing unit.

Acknowledge with

.

„tECH“ „Unit“

CFS/CCS-BA-e-1422

61