14.3 Model KERN CFS 50K-3

Operation

Display

 Switch-on balance and during the selftest press

.

„Pin“

 Use the numeric keys to enter password:
 Either input four times zero „0000“ as a standard password,

or input the user-defined password (input cf. parameter
„Pin“ chapter 13).

 Confirm input by

.

„Pin“ „----„

 Select bulk material or reference balance via

. The

appeared [

] displays the current balance.


When using as counting system, the bulk material balance
as well as the reference balance must be adjusted. The
adjustment process must be carried out on both balances.

 Acknowledge with

.

„tECH“ „LoCAL“

„tECH“ „rEmotE“

 Using select the weighing unit [kg or lb] which will be used

for linearization. The appeared [

] displays the current

weighing unit.

Acknowledge with

.

„tECH“ „Unit“

 Ensure that there are no objects on the weighing pan.

 Wait for stability display (indicator [] atop

appears), then press

.

 Carefully place required adjustment weight (chapter 1) into

the center of the weighing plate on display „LoAd“.

 Wait for stability display, then press

.

 After the adjustment the balance will carry out a self-test.

Remove adjusting weight during selftest, balance will return
into weighing mode automatically.
In case of an adjustment error or incorrect adjusting weight
the display will show an error message (

), repeat

adjustment process.