15 Linearization


Linearity shows the greatest deviation of a weight display on the scale to the value
of the respective test weight according to plus and minus over the entire weighing
range.
If linearity deviation is discovered during a monitoring of test resources, you can
improve this by means of linearization.

• Carrying out linearization is restricted to specialist staff possessing well

acquainted with the workings of weighing scales.

• The adjustment weights to be used must be adapted to the weighing scale’s

specifications; see chapter 3.4 “testing instruments control”.

• Provide for required adjustment weights; cf. subsequent table 1 or table 2.

• Observe stable environmental conditions. Stabilisation requires a certain

warm-up time.

• After successful linearisation you will have to carry out calibration;

see chapter 3.4 “testing instruments control”.

Access to menu:

 Switch-on balance and during the selftest press

.

 Use the numeric keys to input password „9999“ on display „Pin“

 Confirm input by

.



Table 1:
Required adjustment weights KERN CFS

Max

1.

2.

3.

4.

300 g

50 g

100 g

200 g

300 g

3 kg

0.5 kg

1 kg

2 kg

3 kg

6 kg

2 kg

6 kg

-

-

15 kg

5 kg

15 kg

-

-

30 kg

10 kg

30 kg

-

-

50 kg

15 kg

30 kg

50 kg

-