3 remote control instructions, 1 all models except for kern cfs 50k-3, 2 control commands – KERN & SOHN CFS 50K-3 User Manual

Page 79: 3 printing commands, Remote control instructions, All models except for kern cfs 50k-3, Control commands, Printing commands

Advertising
background image

17.3 Remote control instructions

17.3.1 All models except for KERN CFS 50K-3

All inputs finish with <CR><LF> (carriage return / line feed).
At wrong inputs the command will be preceded by "ER", e.g. order "NN<CR><LF>",
error message "ER NN<CR><LF>".

17.3.2 Control commands

PLU

xx

Call article from data storage

T

Tare placed weighing vessel

T123.456

Use the numeric keys to input tare value e.g. 123.456

Z

Zeroing

P

Printing

M+

Add and print weighing data in the summation memory

MR

Call data from sum storage

MC

Delete total added memory

U123.456

Use the numeric keys to input the average article weight
123.456 [g] or [lb].

S123

Define average article weight via weighing.

Function identical to

key.

SL

Switch over to reference balance

SR

Switch over to bulk scales

17.3.3 Printing commands

\L

Selection reference or bulk material balance

\I

User identification number

\S

Balance identification number

\N

Net weight

\G

Gross weight

\U

Average piece weight

T

Tare value

\P

Counting

\C

Total number of pieces

\W

Total weight

\M

Number of summing processes

\B

Insert space line

CFS/CCS-BA-e-1422

79

Advertising