4 model cfs 50k-3, 6 register / call article via rs232, Model cfs 50k-3 – KERN & SOHN CFS 50K-3 User Manual

Page 80: Register / call article via rs232

Advertising
background image

17.3.4 Model CFS 50K-3

Do not finish inputs with <CR><LF> (carriage return / line feed).

T

Taring

Z

Zeroing

P

Printing

17.3.5 User identification, scale identification, store article name

SUID

xxxxxx

<CR>

User identification number
no more than 6 characters

SSID

xxxxxx

<CR>

Scale identification number
no more than 6 characters

SSID

xx,

xxxxxxxxxxxx

<CR>


Storage place
2 characters +
comma

Article name
No more than 12 characters

Unavailable in model CFS 50K-3

17.3.6 Register / call article via RS232

Register article:

Function

Command

1.

Input tare value e.g. 500 g

T0.500<CR>

Where no tare value is needed, input zero

T0<CR>

2.

Average piece weight
e.g. enter 12.3456 g/ piece

U12.3456<CR>

3.

Storage place e.g. 1 (PLU01) followed by the
article name, e.g. M4 screws

SPLU01,M4screws<CR>


Call article:
Command „PLUxx <CR>„, e.g. „PLU01“:
The stored tare value, e.g. 500 g, the average piece weight, e.g. 12.3456 g and the
article name, e.g. „M4 screws“, will be called and displayed.

Unavailable in model CFS 50K-3