1 error messages, Error messages – KERN & SOHN CFS 50K-3 User Manual

Page 84

Advertising
background image

19.1 Error messages

Error message

Description

Possible reasons/

solution

Err 4

Zeroing range exceeded
due to switching-on balance

or pressing

(normally

4% max)

• Object on the weighing

plate

• Overload when zeroing

• Inappropriate

adjustment

• Damaged weighing cell

• Damaged electronics

Err 5

Keyboard error

• Unpurposeful

operation of the scale

Err 6

Value outside the A/D
changer range

• Weighing plate not

installed

• Damaged weighing cell

• Damaged electronics

Err 19

Zero point displaced

• Remedy:

Adjust / linearize

FAIL H / FAIL L

Adjustment error

• Inappropriate

adjustment


Should other error messages occur, switch balance off and then on again. If the
error message remains inform manufacturer.

84

CFS/CCS-BA-e-1422

Advertising