KERN & SOHN KFS-TM User Manual

Page 16

Advertising
background image

 Acknowledge by

and select desired setting with

.

= Linearization

= Adjustment

How to carry out adjustment:

 Confirm menu setting nonLin with

.

Ensure that there are no objects on the weighing pan.

LoAd will be displayed after standstill control has been carried out.

 Put the required adjustment weight carefully in the centre of the weighing pan.

 After successful adjustment, the weighing scales will carry out a selftest.

During this selftest remove the adjustment weight and the weighing scales will
automatically return to weighing mode.
An adjusting error or incorrect adjusting weight will be indicated by the error
message; repeat adjustment procedure.

16

KFS-TM-BA_IA-e-1410

Advertising
This manual is related to the following products: