KERN & SOHN KFS-TM User Manual

Page 42

Advertising
background image

Overview:

Menu item

Available settings

F0 SEL

Enable tolerance check

1 SEL0

Tolerance control disabled

1 SEL1

Tolerance control for weighing

1 SEL2*

Tolerance control for counting

F1 Co

Display conditions of the tolerance
marker

11 Co0

Tolerance marker is always displayed, even if
standstill control is not yet displayed.

11 Co 1*

Tolerance marker is only displayed in connection
with standstill control.

F2 Li

Tolerance range

12 Li 0

Tolerance marker is only displayed above zero
range.

12 Li 1*

Tolerance marker is displayed for the whole
range.

F3 Pn

Number of limiting points

13 Pn 0

1- Limiting point (OK/ -)

13 Pn 1*

2- Limiting points (+/OK/-)

F4 bU

Audio signal

14 bu0*

Audio sound during tolerance control disabled

14 bu1

Audio sound when load is within tolerance limits

14 bu2

Audio sound when load is beyond tolerance limits

F5 Ao

Automatic zero point correction
(zero tracking)

2 Ao0

Automatic zero tracking off

2 Ao1

Automatic limiting point correction on, 0.5 d

2 Ao2*

Automatic limiting point correction on, 1 d

2 Ao3

Automatic limiting point correction on, 2 d

2 Ao4

Automatic limiting point correction on, 4 d

F6 At

Automatic cutout

on

Automatic shut-off ON

off

Automatic shut-off OFF

F7 AP

Automatic shutdown for battery
operation

3 Ap0*

AUTO OFF function disabled

3 Ap1

Instrument will be switched off after 3 minutes of
inactivity of display unit or weighing bridge.

F8 UA

RS-232 mode

4 UA0

Output via RS232C interface disabled

4 UA1*

Continuous data output

4 UA2

Continuous data output of stable weighing values

4 UA3

One output for stable weighing value. No output
for stable weighing values. Renewed output after
stabilization.

4 UA4

For remote commands,

see chap. 9.2.

Issue after pressing the PRINT ´button

4 UA5

Standard printer setting, output after pressing the
PRINT button
id on/off

Printout memory on/off

dt on/off

Printout date on/off

G on/off

Printout gross weight on/off

n on/off

Printout net weight on/off

C on/off

Printout total on/off

PCS on(off)

Printout parts counting on/off

4 UA6

Not documented

42

KFS-TM-BA_IA-e-1410

Advertising