Dual range balance

Confirm with

, then the following menu items can be

selected by

.

dESC

Position decimal point available
selection 0, 0.0, 0.00, 0.000, 0.0000,
0.00000

inC

div 1

inC 1

Readability for

1. Weighing range

Selectable 1, 2, 5, 10, 20,
50

inC 2

inC 5

inC 10

inC 20

inC 50

div 2

inC 1

Readability for

2. Weighing range

Selectable 1, 2, 5, 10, 20,
50

inC 2

inC 5

inC 10

inC 20

inC 50

CAP

CAP 1

Balance capacity (Max) 1st weighing
range

CAP 2

Balance capacity (Max) 2nd weighing
range

Adjust weighing system after configuration.

CAL

nonLin

Adjustment, see chap. 6.5

LinEAr

For linearisation see chapter 6.6