7.2 Menu settings

Menu item „P1 ADJUSTMENT“

[Adjustment]

ILJ-C

ILS-C

P1.1 INT.JUST.

-

Internal adjustment, ILJ-C only

-

P1.2

EXT.JUST.

External adjustment *

-

P1.3

BE.JUST.

User adjustment

P1.4 TEST JUST.

-

Adjustment test

* = Locked in weighing scales with type approval certificate [Er 9 lock].


7.2.1 P1.1 INT.JUST.

– Internal adjustment (models ILJ-C only)

For automatic internal adjustment see chap. 7.1

Starting internal adjustment manually

Press the CAL-

key or invoke menu “P1.1 INT.JUST.“

Press

or

Internal adjustment will be taking place.

Ensure that there are no objects on the weighing plate.

You can cancel the adjustment by pressing the UNIT-key.

After successful adjustment the weighing scale will return automatically to
the menu item. To return to weighing mode, press UNIT.