1. TWO-POINT ADJUSTMENT

The scales must be turned on and you have already carried out a reset to zero.

ƒ Now, press the

PRE-

TARE

key for approximately 20 seconds.

The scales show the value for the first adjustment point, the NET display
arrow is flashing.

Use the

PRE-

TARE

,

TOTAL

and

TARE

function keys to enter the existing adjustment

weight as follows:

ƒ Briefly press the

TARE

key; the right digit starts flashing.

ƒ Use

PRE-

TARE

or

TOTAL

key to set this digit according to the first adjustment weight, e.

g. for 1472 kg, set "2".

ƒ To confirm, briefly press the

TARE

key; now the second digit is flashing.

ƒ Set these to „7“ using

PRE-

TARE

or

TOTAL

.

ƒ To confirm, briefly press the

TARE

key; now the third digit is flashing.

ƒ Use

PRE-

TARE

or

TOTAL

to set this digit to “4”, the hundreds digit to “4“, the thousands

digit might have to be set to “1“ as required, as before the first digit. To

confirm all segments, press the

TARE

key until only the NET display arrow

flashes.

ƒ Import the adjustment weight (in this case the 1472 kg above).

ƒ To confirm, press the

TARE

key for c. 4 sec. The display returns to “AF00“.

Display shows again the first adjustment value.

ƒ Press

the

TOTAL

key, the

~ display arrow is flashing; for a two-point adjustment,

the medium adjustment value must be set to “00000“.

ƒ Briefly press the

TARE

key; the right digit starts flashing.

ƒ Press the function keys

PRE-

TARE

,

TOTAL

and

TARE

to enter the medium adjustment

value (00000 kg) as described above , press the

TARE

key to confirm all

segments until only the

~ display arrow flashes.

ƒ To confirm, press the

TARE

key for c. 4 sec. The display reverts to “AF00“. The

display shows again the medium adjustment value.

ƒ Press

the

TOTAL

key, NET +

~ display arrows are flashing; for a two-point

adjustment the last adjustment value must be set to “00000“.

ƒ Briefly press the

TARE

key; the right digit starts flashing.