Press the function keys, Key; the right digit starts flashing. use, Key; now the second digit is flashing. ƒ use – KERN & SOHN VHS 2T1 Addendum User Manual

Page 3: Key; the third digit is flashing. ƒ use

Advertising
background image

ƒ Press the function keys

PRE-

TARE

,

TOTAL

and

TARE

to enter the last adjustment value,

too, at “0000.0“, as described above, press the

TARE

key to confirm all

segments until only the NET +

~ display arrows flash.

ƒ To confirm, press the

TARE

key for c. 4 sec. The display reverts to “AF00“.

Again the display shows the last adjustment value.

ƒ To exit adjustment mode, press the

TOTAL

key; the scales show e. g. AP 11.

(internal value)

ƒ Press

the

TARE

key for c. 4 sec.; system reverts to weighing mode. On the

display appears the current weight placed on the forks. Adjustment is
complete and you may now take off the adjustment load.

or:

2. MULTIPLE-POINT ADJUSTMENT

The scales must be turned on and you have already carried out a reset to zero.

ƒ Now, press the

PRE-

TARE

key for approximately 20 seconds. The scales show the

value of the first adjustment point; the NET display arrow is flashing.

Use the

PRE-

TARE

,

TOTAL

and

TARE

function keys to enter the existing adjustment

weight as follows:

ƒ Briefly press the

TARE

key; the right digit starts flashing.

Use

PRE-

TARE

or

TOTAL

to set this digit according to the first adjustment weight, e. g.

for 250 kg, set "2".

ƒ To confirm, briefly press the

TARE

key; now the second digit is flashing.

ƒ Use

PRE-

TARE

or

TOTAL

to set this digit to “5”.

ƒ To confirm, briefly press the

TARE

key; the third digit is flashing.

ƒ Use

PRE-

TARE

or

TOTAL

to set this to “2“, the hundreds digit at “2“ and, possibly the

thousands digit at “0“, as required. To confirm, press the

TARE

key until only the

NET display arrow flashes.

ƒ Import the adjustment weight (in this case the 250 kg above).

ƒ To confirm press the

TARE

key for c. 4 sec. The display reverts to “AF00“. The

display shows again the first adjustment value. The first adjustment load (250
kg) can now be taken off.

ƒ Press

the

TOTAL

key, the ~ display arrow is flashing, the medium adjustment

value can be entered.

KERN VHB_VHS

3

Advertising
This manual is related to the following products: