9 error messages – KERN & SOHN HFT 15T5 User Manual

Page 37

Advertising
background image

HFT-BA-e-1310

37

9 Error messages


Error
message

Description

Possible causes

--ol-

Maximum load exceeded

Reduce load

Check whether the balance has
been damaged

Err 4

Zeroing range exceeded due
to switching-on balance or

pressing

(normally 4%

max)

Load attached

Overload during setting to zero

Err 5

Keyboard error

Improper operation of the balance

Err 6

Value outside the A/D changer
range

Damaged weighing cell

Damaged electronics

Err 10

No signal

Distance emitter / receiver too
long

Emitter (connecting box) not
switched on

Lo bA

Capacity of rechargeable
battery exhausted

Recharge battery


Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.





Advertising