10 Menu

Access to service menu „tCH“is locked in verified balances.
To disable the access lock, destroy the seal and actuate the
adjustment switch. For position of adjustment switch, see
chap. 13.

Attention:
After destruction of the seal the weighing system must be re-
verified by an authorised agency and a new verification
wire/seal mark fitted before it can be reused for applications
subject to verification.

10.1 Navigation in the menu

Call up menu

 Turn on the scale during the self-test press

, the first

function [F1 oFF] is displayed.

Select function

 With help of

, the individual functions can be selected

one after the other.

Change settings

 Confirm selected function by

. The current setting will

be displayed.

 Select the desired setting with

and press

to

confirm or

to reject, the balance returns to the menu.

Exit menu/

Return to
weighing mode

 Press

, the balance will return to weighing mode.