MBC-M-BA-e-1312

 Ensure that there are no objects on the weighing pan.

 Wait for stability display "STABLE", then confirm with

(example)

 The size of the currently set adjustment weight is displayed,

the active site flashes.

If required, select with

he digit to be altered and change

the digit with

.

Confirm with

, [LoAd] appears

 Put the required adjustment weight carefully in the centre of

the weighing pan.

Wait until stability display „STABLE“ appears

 Confirm with

[PASS] is displayed.

After the adjustment the balance will carry out a self-test. Remove
adjusting weight during selftest, balance will return into weighing
mode automatically.
An adjusting error or incorrect adjusting weight will be indicated
by the error message; repeat adjustment procedure.

An adjusting error or incorrect adjustment weight will generate an
error message („Err 4“), repeat the adjustment process.