13 Instant help

In case of a fault in the program sequence, the balance should be shortly switched
off. The weighing process must then be restarted from the beginning.

Failure:

Possible causes:

The displayed weight
does not glow.

 The balance is not switched on.
 The mains supply connection has been interrupted

(mains cable not plugged in/faulty).

 Power supply interrupted.
 Rechargeable battery inserted incorrectly or empty
 No rechargeable battery inserted

The displayed weight is
permanently changing

 Draught/air movement
 Table/floor vibrations
 The weighing pan is in contact with foreign bodies

or is not correctly positioned.

 Electromagnetic fields / static charging (choose

different location/switch off interfering device if
possible)

The weighing result is
obviously incorrect

 The display of the balance is not at zero.
 Adjustment is no longer correct.
 Great fluctuations in temperature.
 The balance is on an uneven surface.
 Electromagnetic fields / static charging (choose

different location/switch off interfering device if
possible)

Should other error messages occur, switch balance off and then on again. If the
error message remains inform manufacturer.