5.6.1 Procedure when adjusting (only dealers authorised by KERN)

Observe stable environmental conditions. A warming up time (see chapter 1) is re-
quired for stabilization.

Operation

Display

Turn on balance by pressing the [ON/OFF] key

Keep [

0

] button depressed for approx. 3 sec until

"SETUP" is displayed followed by "9rAd"

Press the [ TARE ] button as often as required until
"CAL ib" is displayed

Press the [HOLD] button.

Press the [TARE] key. ( Zero count mode )

Press the [

0

] button repeatedly until "CAL 0" is

displayed.

Press the [HOLD] button.

Press the [TARE] key.

Enter the required size of the adjustment weight (See
Section 1, "Technical Data"): Press the [HOLD] button
to select the item to be changed and press the [TARE]
button to select the numerical value.

Confirm by pressing the [

0

] button.