10 Instant help


In case of an error in the program process, briefly turn off the balance and discon-
nect from power supply. The weighing process must then be restarted from the be-
ginning.

Help:

Fault

Possible cause

The displayed weight does
not glow.

The balance is not switched on.

The mains supply connection has been interrupted

(mains cable not plugged in/faulty).

Power supply interrupted.

(Rechargeable) batteries are inserted incorrectly or

empty

No (rechargeable) batteries inserted.

The displayed weight is
permanently changing

Draught/air movement

Table/floor vibrations

The weighing plate is in contact with foreign bodies

or is not correctly positioned.

Electromagnetic fields / static charging (choose

different location/switch off interfering device if
possible)

The weighing result is ob-
viously incorrect

The display of the balance is not at zero

Adjustment is no longer correct.

Great fluctuations in temperature.
Warm-up time was ignored.
Electromagnetic fields / static charging (choose

different location/switch off interfering device if
possible)





Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.