MPD_M-BA-e-1312

6.2 Electromagnetic interferences

Guidelines and

manufacturer’s declaration – electromagnetic interferences

The personal scale MPD-M is designed for use in an electromagnetic environment that meets the
requirements stated below. The customer or user of the medical electrical device must ensure that
operation takes place in such an environment.

Emitted interference
measurements

Conformity

Electromagnetic environment -
guideline

HF emissions

as per CISPR 11 / EN 55011

Group 1

The personal scale MPD-M uses HF
energy merely for its internal working.
Its HF emission therefore is very low
and it is highly unlike to interfere with
adjacent electronic devices.

HF emissions

as per CISPR 11 / EN 55011

Class B

The personal scale MPD-M is
designed for use in all equipment
including those in living areas and
those connected directly to the public
power grid that also supplies buildings
used for living purposes.

Emission of harmonics

as per IEC 61000-3-2

Class A

Emission of voltage fluctuations /
flicker

as per IEC 61000-3-3

Conforms with

Do not put the personal scale MPD-M directly next to other devices and do not stack
it with other devices. If this type of operation is necessary, observe the personal
scale MPD-M to ensure normal operation in such an arrangement.