12 Instant help


In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.

Fault

Possible cause

The displayed weight does
not glow.

The balance is not switched on.

The mains supply connection has been interrupted

(mains cable not plugged in/faulty).

Power supply interrupted.

Rechargeable battery inserted incorrectly or empty

No rechargeable battery inserted

The

displayed

weight

is

permanently changing

Draught/air movement

Table/floor vibrations

The weighing plate is in contact with foreign

bodies or is not correctly positioned.

Electromagnetic fields / static charging (choose

different location/switch off interfering device if
possible)

The

weighing

result

is

obviously incorrect

The display of the balance is not at zero

Adjustment is no longer correct.

Great fluctuations in temperature.

Warm-up time was ignored.

Electromagnetic fields / static charging (choose

different location/switch off interfering device if
possible)


Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.