Position adjustment switch and seals:



1.

Self-destroying seal mark

2.

Cover

3.

Adjustment switch

13.1 Verification validity period (current status 2012 in G)

Personal scales (including chair and wheelchair scales) in
hospitals

4 year

Personal scales, when not located in hospitals (for example,
doctor's offices and nursing homes)

unlimited

Baby weighing scales and mechanical birth weight scales

4 year

Bed scales

2 year

Scales in dialysis stations

unlimited


Rehab clinics and health authorities are treated as hospitals.
(4 years of verification validity)

Not treated as hospitals (verification validity not limited) are dialysis stations, nursing
homes and doctor‟s surgeries.

(Details derived from: „Information by the verification authority, weighing scales
applied in medical use“)