16 verification, Z000, Setup – KERN & SOHN MCB 300K100M User Manual

Page 45: Unkt, Cal u

Advertising
background image

16 Verification


Verified scales bear a verification mark or one or more seals affixed by the Bureau of
Standards or the manufacturer on or inside the housing which will self-destroy on
removal. This makes it impossible to verify scales without damaging the seals.

16.1 Adjustment

Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization.

In calibrated balances the adjustment function is switch locked.
In order to carry out adjustments, the switch must be turned to
adjustment position (centre position). (see chapter 15.2).



Display

Operation

a

z000

��

 Start balance by pressing

OFF

ON

SETUP

��

UNKT

��

 Press

for approx. 3 sec,

in the display appears „SETUP“, followed by „UNIT“

CAL

��

K

B

 Press

TARE

repeatedly until “CAL iB“ appears

CAL U

��

 Press

HOLD

, „CAL U“ appears

 Press

TARE

, the appeared triangle

 must be located in the

upper left part of the display.

If not, press again

TARE

.

MCB_MPT-BA-e-1418

45