16.5 Verification validity period (current status in G)
Personal weighing scales in hospitals
4 years
Personal weighing scales if placed outside hospitals
without time
Baby and mechanical scales Infant scales
4 years
Bed scales
2 years
Wheelchair scales
2 years
The hospitals also include rehabilitation clinics and health centres
(4-year validity of verification).
The hospitals do not include dialysis centres, care homes and consultation rooms
(verification validity without time limit).
(Data on the basis: „Verification office informs, scales in medical applications”)