8. CE Declaration of conformity























FH Adjustment Procedure

1 Start the

instrument

Press ON/OFF

The green
light is on

2 Enter the

adjustment
procedure

Directly after pressing
ON/OFF, press PEAK
and PRINT together for
more times and very
short intervals
until the

red light is on.

The lower
red light is
on

3 Instrument

Type

Press SET
Directly after pressing
PEAK + PRINT

XXX N is
shown

a (Back to the

normal
mode ???)

(If you are in the
operation mode again,
the instrument has to be
shut off. Start again with
1. Eventually pressing
the buttons faster)

4 Select the

instrument
type

Press

▼▲

The
correspondi
ng value to
the instr.
appears in
the display

5 Save

instrument
type

Press SET

6 Enter the

adjustment
sequence

Press MEMORY

The upper
red light is
on

7 Chose the

adjustment
weight

Press UNIT and
with

▼▲ specify the

calibration weight in
Newton (X kg * 9,81)

The weight
is written in
the display

8 Save the

adjustment
weight

Press SET and UNIT
simultaneously

9 Attach the

weight to the
instrument
(by hanging)

Press ZERO

The
instrument
changes to
operation
mode.
Adjustment
procedure is
finished