❸Enter Calibration interface, as shown in Fig 5-1.

Calibration interface is shown in
Fig 5-2

❹ Load a standard force. Now the value in standard input area is just equal to the
current measured value. Wait a moment for the force stabilization.

❺ Press

and

to input the standard force value.

❻ Press

to enter the next calibration. Pressing

can interrupt the calibration.

When the 3 times calibration had been finished or been interrupted, a confirm win-
dow will pop up for asking you "Save and Exit" (YES) or “Save and Exit” (NO), (Fig 5-
3a)

Press

or

to select, then press

.


If "YES" is selected, "Calibrate complete!" is displayed, see 5-3(b).

5.4 Troubleshooting


If the force gauge in use processes such as errors, please accord to the following
way for troubleshooting, see table 5-3. Do not disassemble the gauge by yourself to
repair. If you cannot resolve the fault yourself, please contact our company.