GW2045 – GW1060 Manual

Pag. 18 di 43

5.6. ELECTRICAL

CONNECTION

Refer to the technical features diagram for electrical absorption in the section "GENERAL
SPECIFICATIONS
"
.

GW2045 - The machine is designed for connection to an electricity main with the following voltage rating:

~ 230V

1 / N / PE

50Hz



GW1060 The machine is designed for connection to an electricity main with the following voltage rating:

~ 400V

3 / N / PE

50Hz

(GW1060)


GW1060 Family - There is also a single-phase and 230V three-phase "without neutral", with the following
electrical characteristics:

~ 230V

1 / N / PE

50Hz

(GW1060-1)

~ 230V

3 / PE

50Hz

(GW1060-3)


IMPORTANT

GW2045 – The machine has a power cable with a Schuko plug for connection to
electricity main: The plug comprises an appliance sectioning device, which must be
easily accessible for the user.
Wherever possible, prepare a dedicated power sectioning device in the system:

1. Il must be easily accessible for the user.
2. It shall be in close proximity to the equipment.
3. It shall be clearly marked as the sectioning device for the equipment, e.g.

GW2045 DISCONNECTING SWITCH.


Gw1060 – The machine has a power cable for fixed connection to an electrical panel. It is
necessary to prepare a power supply sectioning device (disconnecting switch) in the
system.

1. EASILY ACCESSIBLE
2. CLEARLY MARKED AT THE DISCONNECTING DEVICE FOR THE

APPLIANCE.

Look at the product technical features table at the start of the manual for the correct size
of the disconnecting switch.

fig. 7 - picture showing electrical connection on GW1060.

IMPORTANT
It is essential for the electricity mains to which the machine is connected to comply with
the current standards in force (CEI 64-8/7;V2 standards

3

).

In all cases, make sure that the

earth socket is effective.
SMEG declines all liability for damage caused by connection to a defective socket that
fails to ensure a perfect connection to the general earth conductor, or by a poorly efficient
grounding circuit.

3

CEI 64-8;V2: "Impianti elettrici utilizzatori a tensione nominale non superiore a 1000 V in corrente alternata e a 1500 V in corrente

continua."

http://www.ceiweb.it/