SEA Series Steam Sterilizers

7

5. Operation

Please connect the power cord before you start the operation. Press the main power switch at
right lower corner of the front panel. The power switch light turns on indicating that power is
connected. It will indicate

Ld

when pass examination.

5.1 Fill water
If the water level is lower than the lowest water level. When the power switch is on, select the
sterilize program, the

water

indication light may be glinting while alarm Beeps. At this

time, you need to refill the reservoir. Please open the water tank and pour water into it by
container. Please pay attention to the water level during filling water, you should stop filling
when the surface is close to the red water level mark.

Fig 5-1

Caution:
1). Distilled water must be used to prolong life of the machine.
2). Don't tilt the autoclave when there is water in the reservoir.
3). The condenser water collector should be emptied frequently. It is suggested
that the collector be emptied when refilling the water reservoir.
(Please refer to item 9.2 draining the water.)