Т³гж 27, Appendix, Form1 cycle mode chart – LABEC AA-22N User Manual

Page 26: Form 2 work mode chart, Sterilize mode, as following, Operation manual

Advertising
background image

Operation manual

SEA Series Steam Sterilizers

22

Appendix.

Form1 Cycle mode Chart

Sterilize mode, as following

Form 2 Work mode chart

Item

1

2

3

4

5

6

7

8

Digital display

Mode

Description

Ld

Pre-heating

Preparation and conditioning of the

autoclave

HE

Heating

Chamber is being heated and pressurized

Cycle time

Sterilizing

Counting down of sterilizing

Dr_

Venting

Chamber starts to vent and dry

Drying time

Drying

Instruments are being dried.

Ed_

Complete

Cycle is complete and chamber door can

be opened

Faults & Error codes

Error codes

Displays error codes, resets and alarm

LE_

Balance state

When change the state, the pressure of inside

and outside the Autoclave should be the same.

Sterilizing

temperature

Classes

Pre-heating min

Sterilizing time min

Drying time min

Quick drying time

Total time min

Admission time min

Sterilizing pressure

134

121

2.10

1.10

N

N

0-7

0-7

5

5

4

16

10

10

1/5

1/5

26(16/20) 38(29/33)

o

134 C

o

121 C

Clean test

Clean 20

5

5

25

Advertising