LABEC AA9 User Manual

Page 3

Advertising
background image

 

 
 
 
 
 
General 
The LABEC Automatic Autoclave is designed and constructed to meet the requirements of the

 

 

 

   

 

 

   

 

 

   

 

S.A.A. boiler Code and has been approved by the Boiler Inspectors Branch of the N.S.W.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Department of Industrial Relations ‘Workcover Authority’, to operate at pressures up to 138 K.P.A.

   

 

 

 

   

   

 

   

 

 

Regulations require that on installation the autoclave be examined by a licensed boiler inspector

 

 

 

 

 

 

 

 

 

   

 

 

 

and this certificate of inspection be displayed under glass near the autoclave and renewed annually. 
Approval Numbers 

 

NSW 

QLD 

SA 

WA 

TAS 

AA9 

6766­P­87 

 

 

WA94 

 

AA13/AA16 

674­P­90 

BV18716 

PV86636/7 

WA9489 

TA4946 

AA18/AA20 

673­P­90 

BV18717 

PV86636/7 

WA9490 

TA4947 

 
Heating 
Heating of the autoclave is by means of stainless tubular element located in the base of the

 

 

 

   

 

   

 

 

 

   

 

   

 

autoclave vessel. 
 
Description of Controls 
See Appendix A Attached 
 
Temperature Control 
Labec autoclaves are fitted with solid state proportional action temperature controllers which operate

 

 

 

 

 

 

 

 

 

 

 

 

 

from a sensitive thermocouple or Rtd inserted in the working space of the chamber. The controller

   

 

   

 

   

 

 

   

 

 

 

 

has been calibrated at 121°C and before any adjustments to temperature settings are made allow

 

 

   

 

 

 

 

   

 

 

 

 

 

the autoclave temperature to stabilise for at least one half hour. For the digital controller fitted

 

 

 

 

 

   

 

 

 

 

 

 

 

 

 

please read the enclosed operating instructions when setting the controller temperature. Note: if

 

 

 

 

 

 

 

 

 

 

 

   

you intend using the autoclave at temperatures other than the preset 121°C, you need to

 

 

 

 

   

 

 

 

 

 

 

 

   

adjust the ‘AL4’ value, using the up or down arrow, set this value equal to the new

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

temperature at which you use the autoclave (you must do this in LVL2 in the controller).

   

 

 

 

 

 

 

 

 

   

   

 

 

Now change the temperature using up and down arrows to the new temperature setting

 

 

 

 

 

 

 

 

   

 

 

 

 

and press start. (AL4 is the value which when reached will activate the timer). 
 

Important Safety Instructions 
 

Caution 
Please observe the following safety measures before using your LABEC equipment. 
1.

Ensure pressure inside the autoclave is reading zero KPA on the pressure gauge before

 

 

 

 

   

 

 

 

 

 

 

 

 

opening the lid. 

2.

These autoclaves are NOT FLAME PROOF and under no circumstances should

 

 

 

 

 

 

 

 

 

 

 

inflammable, combustible or explosive material be placed in the autoclave. 

3.

Low ignition temperature materials and those materials which give off inflammable or

 

 

 

 

 

 

 

 

 

 

 

 

explosive vapors should not be placed in the autoclave. 

4.

Avoid heating substances which give off corrosive vapor. 

5.

Users are advised of the dangers of heating combustible materials. The manufacturer can

 

 

   

 

   

 

 

 

 

 

 

recommend special types of elements which will prohibit the autoclaves temperature  

6.

Observe those rules pertaining to wiring and installation of electrical appliances as

 

 

 

 

 

 

 

 

 

 

 

 

recommended by the local supply authority. 

 
 
 
 
 
 
 
 

Laboratory Equipment Pty Ltd Instruction Manual ­ Autoclave AA9 Bench Top Model 