Maintenance 
The cabinet is finished with stainless steel and to maintain appearance should be wiped over with a

 

   

 

 

 

 

   

 

 

   

 

 

   

non abrasive cleanser. The chamber is manufactured from stainless steel and may be cleaned with

 

 

 

 

   

 

 

 

 

 

 

 

 

 

a solvent and if grained stainless steel chamber a stainless steel scratch pad may be used. The

 

 

   

 

 

 

   

 

 

 

 

 

 

 

 

pad should be rubbed in the same direction as the grain pattern of the stainless steel. 
The silicone gasket should be cleaned with detergent only, ensuring it is dried completely after

 

 

 

 

 

 

 

 

 

     

 

 

 

washing. To prolong life it may be stored at cool temperatures in a container of water. Gasket

 

 

 

   

 

 

   

 

     

   

 

 

lubricant is available to provide prolonged life and better sealing on the gasket. 
The locking pins should be regularly greased with high temperature grease to reduce binding in the

 

 

 

 

 

 

 

 

 

 

   

 

   

 

guides.  
 
WARNING: THIS AUTOCLAVE SHOULD NEVER BE CONNECTED TO MAINS WATER SUPPLY  
 
 
Troubleshooting 
 
 
 

SYMPTOM 

REMEDY 

No Power 
 
(Indicator Light is off) 
 
 
 
 
 
 

1. Check autoclave is plugged in and power

 

   

   

 

switched on. 
 

2. Ensure mains power supply point is

 

 

 

 

 

 

functioning by using a test appliance on

 

 

   

 

 

 

power socket. 
 

3. Check internal RCD has not tripped. 

 

 

Failure to heat or maintain temperature 
 
(Indicator light is on) 

1. The temperature controller is on. 

 

2. The water level is above the low water

 

 

   

 

 

 

sensor and the low water indicating lamp

 

 

 

 

 

 

is off. 
 

 

The timer does not start 

1. Check the AL4 value is equal to set value

 

 

 

   

   

 

as this value activates the timer. 
 

 

 
If the fault cannot be found call your distributor or the manufacturer quoting the serial number of the

 

 

 

 

 

 

 

 

   

 

 

 

 

 

   

 

autoclave from the manufacturers label. 
 
 
 
Declaration of Conformity 
Each product is thoroughly inspected and tested to not only ensure that it meets the specifications

 

   

 

 

 

   

 

 

 

   

 

 

 

provided, but to also meet Australian Electrical Standard AS3820 and EMC Standard

 

 

 

 

 

 

 

 

 

 

 

 

AS/NZ1044:1995,and therefore being accredited with a C Tick label. Labec Autoclaves are

 

 

 

 

     

 

 

 

 

 

constructed in accordance with Australian Standard 1210 and are approved by Workcover Authority

   

 

 

 

 

 

 

 

   

 

 

of NSW with reciprocal approval by equivalent bodies in all states of Australia. 
 
 

Laboratory Equipment Pty Ltd Instruction Manual ­ Autoclave AA9 Bench Top Model 