LABEC AA9 User Manual

Page 5

Advertising
background image

 

External Steam Fittings 
1.

Spring loaded safety valve pre­set to operate at 138 K.P.A. with padlock. 

2.

Automatic air vent. 

3.

Pressure gauge. 

4.

Solenoid valve & ball valve. 

 
Indicators 

1. An indicating lamp adjacent to the controller will illuminate when the heaters are operating. 
2. An amber pilot light will illuminate if the water level is too low. 
3. An automatic digital reset timer with multiple time modes will be activated once

 

 

 

 

 

 

 

 

 

 

 

 

 

temperature reaches the set temperature, the timer will then flash “off” until the time has

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

elapsed, then the autoclave will cool until the ‘start’ button is pressed again. 

4. If you wish to change the temperature you are using the autoclave at and therefore the

 

 

   

 

 

 

 

 

 

 

   

 

 

 

temperature at which the timer will begin its timing cycle please call the manufacturers

   

 

 

 

 

 

 

 

 

 

 

 

 

technical department. 

5. It is advisable to check the water level after each cycle. 

 
 
Chart Recorder (Optional ­ not on all models) 
Please read the chart recorder manual before use of the autoclave. 
The recorder is programmed to record the autoclave temperature on Channel 1. If the recorder is

 

   

   

 

 

 

 

 

     

 

   

remote it should be mounted/placed adjacent to the autoclave and then using the leads from the

   

 

 

 

   

 

 

 

 

 

 

 

 

 

recorder, connected to the matching sockets on the rear of the autoclave body. When positioned

 

   

 

 

 

 

 

   

 

 

 

 

 

and connected to the autoclave, open the door of the recorder, place the chart paper in position

 

   

 

 

 

 

   

 

 

 

 

 

 

 

 

and remove the white caps from the pens(the internal door can be opened to access the pens),

 

 

 

 

 

 

 

 

 

 

 

 

   

 

 

 

press the button to switch the recorder ON. Connect the recorder to the MAINS supply and turn

 

 

   

 

 

 

 

 

 

   

 

 

 

 

 

ON. 
 
External Steam Fittings 

1. Spring loaded safety valve pre­set to operate at 138 K.P.A. (Safety valve). 
2. Automatic air vent (air vent 2). 
3. Adjustable steam release valve. If autoclaving sealed items, the adjustable steam release

 

 

 

   

 

 

 

 

 

 

 

valve may be closed to prevent the items exploding or blowing off lids or seals when the

 

 

 

   

 

 

 

   

 

 

   

 

 

 

autoclave has completed its cycle. In this case it is desirable to allow the autoclave to

 

 

 

 

   

 

     

   

 

 

   

depressurise gradually before attempting to open the lid. The valve may be partially opened

 

 

 

   

 

 

 

 

 

 

 

 

 

to permit a gradual release of steam (dump valve 5). 

4. Pressure gauge. 

 
 
Internal Temperature Probes (Optional­ not on all models) 
If the autoclave is fitted with two internal temperature probes. To change these probes you must

 

 

   

 

 

 

 

 

 

 

 

 

 

 

 

remove the side panel and unscrew the glands in which they enter the chamber. The probes are ‘K’

 

 

 

 

 

 

 

   

 

 

 

 

 

 

 

 

 

type thermocouple autoclave type and are connected to an external plug for monitoring by chart

 

 

 

 

 

 

   

 

 

 

 

 

 

 

recorder or computer or P.L.C. Included is two blank plugs which you can screw into the vessel if

   

   

 

   

 

 

 

 

 

 

 

 

 

   

you do not wish to use the internal probes­ be sure you use plumbers tape before screwing these

 

 

 

   

 

 

 

 

 

 

 

 

 

 

 

 

 

into the vessel or replacing them in the vessel.  
 
 
 
 
 
 
 
 
 
 
 
 
 

Laboratory Equipment Pty Ltd Instruction Manual ­ Autoclave AA9 Bench Top Model 