Level clear floor surface. 

The ambient temperature of the room where the unit is placed should not exceed 32℃. 

Not subject to mechanical vibration or shock. 

Do not place in high traffic areas. 

Do not leave in direct sunlight. 

Do not place in a dust prone area. 

Do not place unit in a draft prone location. 

Keep away from heat sources such as autoclaves, heating diffusers, radiators, or other

 

 

 

 

 

 

 

 

 

 

 

 

 

sources. 

A minimum 30cm space on the sides, rear and top should be provided for better air circulation,

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

heat discharge and easy of general cleaning duties. 

Make sure that air intakes for compressor cooling are kept free from obstruction.  

 

Notice: After connecting the power supply, unit should be run for at least 12 hours before

 

 

 

 

 

 

 

 

   

 

   

   

 

 

any products are placed in freezer.  

 

 

                                    

 

N and E 

port can’t connect 

2  Safety Introduction 

Power supply: This ULT freezer is designed for operating under 220V/50HZ AC power supply.

 

 

 

 

   

 

 

 

 

 

 

 

 

The unit can normally start at a voltage varying from 198V to 242V. 

Must provide one single exclusive plug receptacle with credible grounding to unit. 

Inflammable and explosive materials as well as high corrosive acid and alkali are strictly

 

 

 

 

 

 

 

 

 

 

 

 

 

 

prohibited in cabinet.  

Keep the key properly avoiding the children take it to open the back door which may result in

 

 

 

 

 

 

 

     

 

 

 

 

 

 

   

unexpected injury. 

To avoid cold injury, please place and remove products in unit with protective gloves.  

 

 

 