Type 2 commands – balance input functions – Scientech 10K Series User Manual

Page 17

Advertising
background image

Command Response

Purpose

[decimal number]_CAL[CR]

None

Loads a computed number based on the entered number, as a
"last front panel units" relative multiplier, into the CAL
register and enters the CAL mode. The display must be non-
zero when this command is issued. If the display is zero, a 3
second "Err" message is displayed and the balance returns to
its normal weighing mode. The current weight reads as the
number entered with the appropriate sign. CLEAR exits the
CAL mode and un-defines the CAL register.

[integer]_PIECES[CR]

None

Loads a computed number based on the entered number, as
a "last front panel units" relative multiplier, into the
PIECES register and enters the PIECES mode. The display
must be non-zero when this command is issued. If the
display is zero, a 3 second "Err" message is displayed and
the balance returns to its normal weighing mode. CLEAR
exits the pieces mode and un-defines the pieces register.

[Decimal number]_ENTER_[integer]
_PIECES[CR]

None

Accepts the decimal number as a weight in "last front panel
units" per integer pieces. If the balance is not in PCS mode,
it goes into PCS mode using the decimal number/the integer
as the weight. If used when the balance is already in PCS
mode, the display digits are blanked for about 1 second
while the PCS and "last front panel units" annunciators are
lit. The balance then resumes normal PCS operation using
the decimal number/the integer as the unit weight.

[register
number]_RCL_[(CONVERT) (CAL)
(PIECES)][CR]

REG XXX :[register
contents][MULT.]

Puts the recalled register contents into the specified mode
register and enters that mode.

TAVG[CR]

None

Clears the TBAR registers and enters the TBAR mode.

RESET_TAVG[CR]

None

Clears the TBAR registers. If the balance is in the TBAR
mode, the process keeps running.

XAVG[CR] None

Accumulates the current weight in the XBAR registers.

RESET_XAVG[CR] None

Clears the XBAR registers.

GLP[CR]

Sends GLP header

Puts balance in the GLP mode.

XAVG[CR]

Sends weight while in GLP

Accumulates the current weight in the XBAR registers.

RCL_XAVG[CR]

Sends statistics and signature
line

Terminates GLP mode.

ISO[CR]

Sends ISO header and waits
for target weight entry

Puts balance in ISO mode.

xxx.xx_Targ[CR]

Sends target weight (g)

Enter target weight value in grams.

SEND[CR]

Sends weight, difference
between weight and target,
and signature line

Terminates ISO mode.

ASON[CR] or ASOFF[CR]

See the RS-232 Balance
Message Formats section.

Turns auto send on or off. When auto send is on, a time
interval must be entered into register 89 (see Type 4
commands).

CSON[CR] or CSOFF[CR]

See the RS-232 Balance
Message Formats section.

Turns continuous send on or off.

ACAL[CR]

None

Starts the automatic internal calibration weight calibration
routine.

Type 2 Commands – Balance Input Functions:

Command Response

Purpose

[register number]_STORE[CR]

None

Stores the current weight, at the greatest possible resolution, in the
specified register.

[Decimal number]_ENTER_[register
number]_STORE[CR]

None

Stores the number entered, in the current units, as an additive
constant in the register specified.

[decimal number]_ENTER_CONVERT_[register
number]_STORE[CR]

None

Stores the number entered, as a "last front panel units" relative
multiplier, in the specified register.

[decimal number]_TARE[CR]

None

Adds the number entered, in current units, to the TARE register.
CLEAR [CR] zeroes the TARE register.

[decimal number]_ENTER_86_STORE[CR]

None

Stores the desired precision of the displayed weight for

T

mode.

17