Troubleshooting Guide:


Note: Many problems such as the balance remaining in the turn on sequence, "weird" characters on the display or

"garbage" on the display are caused by connecting the power supply to the balance incorrectly. The proper
procedure is to connect the round DIN connector on the power supply cable to the balance then plug the power supply
into the wall outlet.

Problem

Possible Cause

Possible Solution

The display is blank

*Balance not turned on
*Power cable not plugged in
*No power from AC outlet

*Press ON/OFF switch
*Press ON/OFF switch

*Turn on circuit breaker
*Change AC outlets

OL appears in display

*Maximum capacity exceeded


*Shipping clamp is engaged

*Reduce container weight

*Weigh sample in smaller increments
*Switch to high range (dual range units)

*Disengage clamp – see page 4

UL appears in display

*Pan is not in place
*Balance is out of range
*Shipping clamp is engaged

*Insure pan is positioned properly
*Press ZERO switch
*Disengage clamp – see page 4

The display is unstable

*Drafts (air currents are present)
*Vibrations are present

*Install windscreen, relocate balance

*Isolate balance from vibration, relocate balance

Weight readings are incorrect

*Balance is out of calibration
*Balance is not level
*Display was not zeroed before weight was
placed on pan
*Object being weighed is touching windscreen
or balance case top
*Unit of weight selection is incorrect

*Recalibrate balance
*Level balance
*Press ZERO switch before weighing

*Reposition object

*Select proper weighing mode

Display stays in turn on sequence *Pan not in place

*Drafts or vibrations present
*New windscreen has been installed
*Shipping clamp is engaged
*Power supply connected to wall outlet before
balance connector

*Place pan on balance
*Isolate or relocate balance
*Raise pan height
*Disengage clamp – see page 4
*Unplug power supply from wall then replug
into wall

Err 1 appears in display

*Calibration weight is incorrect

*Use calibration weight that appears in balance
display

Err 2 appears in display

*Balance not warmed up enough for calibration
to be performed

*Plug in and turn on balance for 1 hour

Err 3 through 14 appears in
display

*Service is required

*Call Scientech's Service Department at (800)
525-0522 or (303)444-1361 or email us at
[email protected]

Err appears in display

*Procedural error

*Review procedure in manual and try again