Control overview, Watlow controller details, Power switch unit – Shellab SVAC9-2 User Manual

Page 13: Power switch vacuum, Over temperature limit thermostat (otl), 13 | p a g e

Advertising
background image

13 |

P a g e

CONTROL OVERVIEW

20.0

20.0

RES E T

EZ1

EZ2

2

C

CURRENT OVEN CHAMBER

TEMPERATURE

TEMPERATURE SET

POINT (YOUR DESIRED

TEMPERATURE)

INDICATES THAT THE

CONTROLLER IS POWERING THE

OVEN’S HEATER

RAMPING SYMBOL INDICATES

THAT A HEATING PROFILE IS

RUNNING

THE ADVANCE KEY SCROLLS

THROUGH MENUS AND

PARAMETER LISTS

THE DOWN ARROW KEY

ADJUSTS SET POINTS OR

CHANGES PARAMETERS

THE UP ARROW KEY

ADJUSTS SET POINTS

OR CHANGES PARAMETERS

THE RESET KEY

SCROLLS 1 PAGE BACK

THE EASY ZONE KEY 1

INITIATES AND ENDS

HEATING PROFILES

Watlow Controller Details

Power Switch Unit

The main power switch on the control panel (green lighted I/O) controls all power to the unit and
must be in the I (ON) position before any systems are operational. The switch will be lighted when
in the on position.

Power Switch Vacuum

This I/O (on / off) switch is marked VACUUM and controls power to a vacuum pump plugged into
the 230VAC outlet in the cabinet. The switch will illuminate when in the on position.

Over Temperature Limit Thermostat (OTL)

Labeled SET OVER TEMPERATURE and equipped with a graduated dial and independent
temperature probe, the OTL system guards against an uncontrolled rise in oven chamber
temperature. This allows continued operation of the oven in the event of a main controller failure,
until the problem can be corrected. Please see the

OTL entry

in the Theory of Operation section on

page 17 for a more detailed description of the system.

Figure 7: Control Panel SVAC9-2 and Watlow Temperature Controller

Advertising