Page 3-2


5. To connect the wires to the terminals, follow the instructions below:

a. Using a small Flat head screwdriver, loosen the terminal screws. This

will open the terminal clamps so that you can insert the wire.

b. Carefully insert the bare end of the wire between the terminal clamps.
c. While holding the wire in position, use the screwdriver to tighten the

terminal screw. The clamps will close around the wire. After it is tight
give the wire a small tug to make sure it’s secure.

d. Repeat the process for the remaining wires.

When inserting the wire into the terminal clamp, make sure the

clamp secures the BARE END of the wire and NOT the colored INSULATION of
the wire.

If the bare end of the wire is not clamped properly the scale will not

operate correctly! This will be the most likely cause of any problem you may
have when using the scale for the first time.

6. Once all the wires are secure, pull out any excess slack in the cable through

the watertight fitting. If there is too much cable inside the box you may not be
able to close the lid.


7. Tighten the liquid tight fitting nut to secure the cable so that moisture can not

enter the junction box.