Appendix a_troubleshoot – Triner Scale TSM5-28-LS User Manual

Page 9

Advertising
background image

A-1

APPENDIX A: TROUBLESHOOTING AND PERIODIC MAINTAINANCE

A.1 Troubleshooting Guide

Problem

Possible Problem and Corrective Action

The system does not
operate – No Display

• Power disconnected or bad power supply: Check outlet

first. Test the AC adapter or power supply.

• Indicator fuse blown: Replace fuse, check for cause.
• Interface Cable cut or disconnected: Replace or Properly

repair cable

• Interface Cable wiring problem: Check connections

inside junction box and at indicator.

Display stays at Zero – Will
not weigh anything

• Item may be too light for scale resolution.
• Interface Cable cut or disconnected: Replace or Properly

repair cable

• Interface Cable wiring problem: Check connections

inside junction box and at indicator.

• Load cell connections faulty: Check load cell connections

inside junction box.

• Debris under the scale platform: Check and clean debris

from under the platform (Periodic Maintenance).

• Indicator faulty. Service or replace indicator.

Displays erratic weights

• Interface Cable cut or disconnected: Replace or Properly

repair cable

• Interface Cable wiring problem: Check connections

inside junction box and at indicator.

• Platform not level: Level the scale by properly adjusting

the leveling feet.

• Load cell connections faulty: Check load cell connections

inside junction box.

• Load cell or load cell cable damage: Inspect load cell

cables. Test and replace Loadcell if needed.

• Bad power supply: Check outlet first. Test the AC

adapter or power supply.

• Debris under the scale platform: Check and clean debris

from under the platform (Periodic Maintenance).

• Indicator faulty. Use simulator to test for stability. Service

or replace indicator.

Consistently high or low
weights

• Indicator not properly adjusted to zero: Zero the indicator

according to the indicator manual.

• Platform binding: Provide adequate clearance for free

platform movement. Make sure a ramp(s) is not touching
the scale. Remove any debris under and around the
scale platform.

• Indicator not calibrated properly: Calibrate the system

according to the indicator manual.

• Load cell(s) faulty: Test and replace load cells if

necessary.