Triner Scale T1000 Error Code User Manual

Page 4

Advertising
background image

10

Program memory EPROM- or customer
EPROM- error

Program memory EPROM or customer EPROM not present, defective, or incorrectly installed. IC- and
socket markings do not agree or IC contact legs make no contact.

Permanent Error code 10 and machine blocks.

If a new cliché EPROM was just installed and the error is: 10 21, 10 22 or 10 29 indicates the EPROM
or its contents do not match this type of machine.

Describe the Error code when ordering a new component.

10

10

Error operating EPROM

Program memory EPROM not present, defective, or incorrectly installed. IC- and socket markings do
not agree or IC contact legs make no contact.
Permanent Error code 10 and machine blocks.

Check components (all pins good contact, correct position, is it the right component?)

10

11

Error operating EPROM
(discovered when calculating value
figures)

Program memory EPROM not present, defective, or incorrectly installed. IC- and socket markings do
not agree or IC contact legs make no contact.
Permanent Error code 10 and machine blocks.

Check components (all pins good contact, correct position, is it the right component?)

10

12

Error operating EPROM
(discovered when calculating date figures)

Program memory EPROM not present, defective, or incorrectly installed. IC- and socket markings do
not agree or IC contact legs make no contact.
Permanent Error code 10 and machine blocks.

Check components (all pins good contact, correct position, is it the right component?)

10

13

Error operating EPROM
(discovered when calculating machine
no.)

Program memory EPROM not present, defective, or incorrectly installed. IC- and socket markings do
not agree or IC contact legs make no contact.
Permanent Error code 10 and machine blocks.

Check components (all pins good contact, correct position, is it the right component?)

10

14

Error operating EPROM
(discovered when calculating counters)

Program memory EPROM not present, defective, or incorrectly installed. IC- and socket markings do
not agree or IC contact legs make no contact.
Permanent Error code 10 and machine blocks.

Check components (all pins good contact, correct position, is it the right component?)

10

19

Error operating EPROM
(discovered when calculating bar codes)

Program memory EPROM not present, defective, or incorrectly installed. IC- and socket markings do
not agree or IC contact legs make no contact.
Permanent Error code 10 and machine blocks.

Check components (all pins good contact, correct position, is it the right component?)

10

1x

Error operating EPROM

Program memory EPROM not present, defective, or incorrectly installed. IC- and socket markings do
not agree or IC contact legs make no contact.
Permanent Error code 10 and machine blocks.

Check components (all legs good contact, correct position, is it the right component?)

diagnostic-
code

subcode1
subcode2

diagnostic message

remedy

ISS international service support
email: [email protected]

created : 20.11.2007 19:31 Uhr

page 4

T1000