Triner Scale T1000 Error Code User Manual

Page 5

Advertising
background image

10

20

Error cliché EPROM

Cliché EPROM not present, defective, or incorrectly installed. IC- and socket markings do not agree or
IC contact legs make no contact.
Permanent Error code 10 and machine blocks.

Check components (all pins good contact, correct position, is it the right component?)

10

21

Error cliché EPROM

Cliché EPROM not present, defective, or incorrectly installed. IC- and socket markings do not agree or
IC contact legs make no contact.
Permanent Error code 10 and machine blocks.

Check components (all pins good contact, correct position, is it the right component?)

If a new cliché EPROM was just installed and the error is:
10 21, 10 22 or 10 29 indicates the EPROM or its contents do not match this type of machine.

10

22

Error cliché EPROM

Cliché EPROM not present, defective, or incorrectly installed. IC- and socket markings do not agree or
IC contact legs make no contact.
Permanent Error code 10 and machine blocks.

Check components (all pins good contact, correct position, is it the right component?)

If a new cliché EPROM was just installed and the error is:
10 21, 10 22 or 10 29 indicates the EPROM or its contents do not match this type of machine.

10

26

Error cliché EPROM

Cliché EPROM not present, defective, or incorrectly installed. IC- and socket markings do not agree or
IC contact legs make no contact.
Permanent Error code 10 and machine blocks.

Check components (all pins good contact, correct position, is it the right component?)

10

27

Error cliché EPROM

Cliché EPROM not present, defective, or incorrectly installed. IC- and socket markings do not agree or
IC contact legs make no contact.
Permanent Error code 10 and machine blocks.

Check components (all pins good contact, correct position, is it the right component?)

10

28

Error cliché EPROM
(wrong customer cliché no)

Cliché EPROM not present, defective, or incorrectly installed. IC- and socket markings do not agree or
IC contact legs make no contact.
Permanent Error code 10 and machine blocks.

Check components (all pins good contact, correct position, is it the right component?)

diagnostic-
code

subcode1
subcode2

diagnostic message

remedy

ISS international service support
email: [email protected]

created : 20.11.2007 19:31 Uhr

page 5