Notice – VACUUBRAND VHC Pro User Manual

Page 17

Advertising
background image

page 17 of 27

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. 25/11/2013 // 999281

Tools required (metric):

- Diaphragm key SW 46 (order no.: 637777)

- Screw driver Torx TX20

- Open-ended wrench SW 19

Decontamination of the pump parts in a disinfectant solution or by autoclaving

The vacuum pump has to be disassembled.

When disassembling avoid contact of contaminated parts with not contami-

nated parts.

The pictured pump parts have to be decontaminated.

The applicability of the decontamination method and the selection of the ap-

propriate disinfectant have to be ensured by the user.

Prior to decontamination clean surfaces with a damp cloth if necessary. Use water or

a mild detergent as cleaning agent.

NOTICE

NOTICE

NOTICE