Decontamination by autoclaving

121°C

2 bar absolute

Autoclaving

The collection bottle with bottle head and screw cap, the quick

coupling, the filter and the VHC

pro

with adapters and aspirati-

on hose are designated for steam sterilization at 121°C and

2 bar absolute (1 bar overpressure). Time of exposure accor-

ding to DIN 58946 t

e

= 20 minutes.

Ensure not to exceed a temperature of 121°C.

Comply with instructions for use of the manufacturer of the

autoclave.

It is the user´s responsibility to ensure effective autoclaving.

Prior to autoclaving loosen or remove the bottle head

from the bottle.

Midi

sart® 2000

0.20

µm PTFE

IN

sartorius stedim

Prior to autoclaving set the VHC

pro

to “continuous aspira-

tion / autoclaving”.

The applicability of the decontamination method has to be ensured by the user.

Prior to decontamination clean surfaces with a damp cloth if necessary. Use water or

a mild detergent as cleaning agent.

NOTICE

NOTICE

NOTICE