Technical data

Type

HP 40B2 /

RZ 2.5

HP 40B2 /

RZ 6

HP 63B2

Backing pump

RZ 2.5

RZ 6

Recommended oil for backing pump

VACUUBRAND B-oil

Diffusion pump

AX 65

Recommended pump fluid for

diffusion pump

diffusion pump oil for

VACUUBRAND oil diffusion pump

Pump fluid capacity of diffusion pump

ml

30

Maximum pumping speed

l/s

22

37

Ultimate vacuum*

mbar

< 1 x 10

-6

Maximum permissible inlet / outlet

pressure (absolute)

bar

1.1

Maximum pressure difference

between inlet and outlet

bar

1.1

Maximum permissible pressure

(absolute) at gas ballast valve

bar

1.2

Maximum permissible temperature

of gaseous media

continuous operation: 40°C,

for short periods (less than 5 minutes): up to 80°C

Permissible ambient temperature

storage / operation

°C

-10 to +60 / +12 to +40

Permissible relative atmospheric

moisture during operation

%

30 to 85 (no condensation)

Maximum permissible installation

altitude above mean sea level

m

2000

Maximum permissible range of

supply voltage

230V~

±10% 50/60Hz

Rated input current:

230 V~ 50/60 Hz

A

2.7 / 2.2

3.2 / 3.2

Power draw

230 V~ 50/60 Hz

VA

621 / 506

736 / 736

Degree of protection IEC 529

IP 10B

Inlet

small flange KF DN 40

clamping flange

ISO-K DN 63

Outlet

hose nozzle DN 10 mm

A-weighted emission sound pressure

level** (uncertainty K

pA

: 3 dB(A))

dB(A)

42

49

Dimensions L x W x H approx.

mm

445 x 385 x 435

445 x 381 x 460

Weight approx.

kg

25.4

30.9

Advertising
This manual is related to the following products: