Ambient conditions, Operating conditions, Notice – VACUUBRAND HP 63B2 User Manual

Page 6

Advertising
background image

page 6 of 20

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999069 / 24/09/2014

Make sure ventilation is adequate to maintain recommended operating temperature.

Keep a minimum distance of 5 cm between the cooling fans and surrounding items

(e.g., housing, walls, etc.), or else install an external automatic ventilation system.

Check fans regularly for dust/dirt. Clean fan guard grills if necessary to avoid a re-

duction of ventilation.

Use only hoses at the inlet and outlet of the pump with an inner diameter at least as

large as the diameter of the pump’s tubing (to avoid overpressure at the outlet, and

reduction of pumping speed at the inlet).

Allow the equipment to equilibrate to ambient temperature if you bring it from cold

environment into a room prior to operation. Notice if there is water condensation on

cold surfaces.

The ON/OFF-switches are located at the terminal box. At first, switch on the rotary

vane pump (backing pump), then switch on the diffusion pump. The diffusion pump

can be switched on only if the backing pump has been switched on.

Comply with all

applicable and relevant safety requirements (regulations and

guidelines).

Implement the required actions and adopt suitable safety mea-

sures.

Ambient conditions

• Adopt suitable measures in case of differences from recommended conditions,

e.g., using the equipment outdoors, installation in higher altitudes, conductive

pollution or external condensation on the pump.

• Do not operate this product near flames.

To the best of our knowledge the equipment is in compliance with the requirements

of the applicable EC-directives and harmonized standards (see ”Declaration of Con-

formity”) with regard to design, type and model. Directive EN 61010-1 gives in detail

the conditions under which the equipment can be operated safely (see also IP de-

gree of protection).

Operating conditions

These pumps are not approved for operation in potentially explosive atmos-

pheres. Do not operate the pumps in potentially explosive atmospheres.

These pumps are not approved for the pumping of potentially explosive

atmospheres. Do not pump potentially explosive atmospheres.

The pumps are

not suitable to pump any of the substances listed below.

Do not pump:

-

unstable substances

- substances which react explosively under

impact (mechanical stress) without air

- substances which react explosively when being exposed to

elevated tempera-

tures without air,

- substances subject to

auto-ignition,

- substances which are

inflammable without air

-

explosive substances.

The pumps are

not approved for operation below ground. Do not operate the

pump

below ground.

NOTICE

NOTICE

Advertising
This manual is related to the following products: