VACUUBRAND MZ 2C +AK+M+D User Manual

Page 12

Advertising
background image

page 12 of 26

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. Manual-no.: 999058 / 05/10/2009

Attention: Important notes regarding the use of gas ballast

☞ Make sure that air/gas inlet through the gas ballast valve never lead to hazardous,

explosive or otherwise dangerous mixtures. If in doubt, use inert gas.

☞ When using air rather than inert gas, risk of significant damage to equipment and/or

facilities, risk of personal injury or even loss of life exists due to the formation of
hazardous and/or explosive mixtures if air and pumped media react inside or at the
outlet of the pump.

During operation:

Do not start pump if pressure difference between inlet and outlet port exceeds maxi-
mum 1 bar
. Attempts to start pump at higher difference may cause blockade and
damage of the motor.
☞ Check compatibility with maximum permissible pressure at outlet and maxi-

mum pressure difference between inlet and outlet ports.

Due to the high compression ratio of the pumps, pressure at the outlet port might be
generated being higher than the max. permitted pressure compatible with the mechanical
stability
of the system.

The pump achieves its pumping speed, ultimate total vacuum and vapour pumping
rate only at operating temperature (after approx. 15 minutes).
☞ Prevent internal condensation, transfer of liquids or dust. The diaphragm and valves

will be damaged, if liquids are pumped in significant amounts.

☞ Let the pump run with gas ballast to reduce condensation of pumped substances

(water vapour, solvents, ....) in the pump.

Motor is shut down by a thermal cutout in the winding.
☞ Manual reset is necessary. Switch off the pump or isolate the equipment from mains.

Wait approx. five minutes before restarting the pump. Identify cause of failure and
eliminate.

Attention: In case of supply voltage below 100V, the lock of the cutout might be

restricted and the pump might restart on its own after sufficient cooling down. Take
suitable precautions, if an automatic restart of the pump may lead to a critical
dangerous situation.

☞ Secure coolant hoses at the hose nozzles (e.g. with hose clip) to prevent their

accidental slipping.

☞ Ensure that the coolant outlet pipeline is always free and that it cannot get blocked.

☞ The gas outlet (hose nozzle 10 mm) must not be blocked. The exhaust pipeline has

always to be free and pressureless to enable an unhindered discharge of gases.

☞ Connect the exhaust to a suitable treatment plant to prevent the discharge of dan-

gerous gases and vapours to the surrounding atmosphere.

For condensable vapours (water vapour, solvents, ...):
☞ Do not pump vapour before pump has reached its operating

temperature and with gas ballast valve closed.

☞ Open gas ballast valve. (Closing of gas ballast valve by turn-

ing 180°.)

☞ The gas ballast valve is open if the arrow on the gas ballast

cap shows away from the pump.

☞ With gas ballast valve open ultimate vacuum will be reduced,

pumping speed is decreased.

☞ Use inert gas at the air inlet to avoid the formation of explo-

sive mixtures.

gas ballast
(open)