If the pump does not achieve the ultimate pressure:

☞ In case the diaphragms and valves have been replaced, a run-in period of several

hours is required before the pump achieves its ultimate vacuum.

If all pump heads achieve a vacuum below 120 mbar but pump does not achieve the
ultimate total pressure:
Check hose connectors between pump heads and manifolds for leaks. If necessary
recheck pump chamber.

➨ Use open ended wrench (w/f 14) to reconnect hose to elbow

fitting.

➨ Tighten union nuts first by hand and then tighten one full

turn using open ended wrench (w/f 17).

Assembling fittings: