Start: Press key START/STOP.

☞ Arrow down and valve and/or pump symbol are flashing: Su-

pervise process permanently.

➨ When sufficient evaporation appears: Press key MODE.

☞ Continuous pumping is stopped. The actual pressure p is

stored as new set point.

☞ Two-point control is activated with set point p and Δp as

preprogrammed on the controller. Hysteresis

Δp can be

changed manually at any time.

MODE

START

STOP

Changing the set point p downwards by using key p

▼:

➨ When the key is pressed shortly, the current pressure is

displayed for one second.

☞ Continuous pressing or a second tip within one second: The

in-line valve is opened, pump is switched on, the current
pressure is displayed and stored as new set point when the
key is released.

☞ Hysteresis Δp is adapted automatically only in case of great

set point changes. Hysteresis

Δp can be changed manually

at any time.

Changing the set point p upwards by using key p

▲:

➨ When the key is pressed shortly, the current pressure is

displayed for one second.

☞ Continuous pressing or a second tip within one second: The

venting valve is opened, the current pressure is displayed
and stored as new set point when the key is released.

☞ Hysteresis Δp is adapted automatically only in case of great

set point changes. Hysteresis

Δp can be changed manually

at any time.

Changing the set point p during process control

p

520

mbar

p

67

mbar