VACUUBRAND MD 1 VARIO-SP User Manual

Page 35

Advertising
background image

page 35 of 68

Changing the selection at the voltage se-

lection switch:

Change the selection at the voltage se-

lection switch only, if the pump is un-

plugged from the power source.

1.

Disconnect the electrical power cord.

2. Use a screw driver to adjust the voltage se-

lection switch to the supply voltage:

”115/120” corresponds to 100-120 V and

”230/240” corresponds to 200-230 V.

MD 1 VARIO-SP (DC motor)

• Check the power source and the pump’s

rating plate to be sure that the power source

and the equipment match in voltage, phase,

and frequency.

Connect the supply voltage and the control

signal to the pump’s control line (see „Control-

ling the MD 1 VARIO-SP“, pg. 36).

Switch the supply voltage and, if required, the

control signal on.

Advertising
This manual is related to the following products: